MAUDE Adverse Event Report: DePuy Orthopaedics Inc US SUMMIT; Summit Hip Stem : Hip Femoral Stem

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DePuy Orthopaedics Inc US SUMMIT; Summit Hip Stem : Hip Femoral Stem Back to Search Results
Catalog Number 157013090
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Osteolysis (2377); Bone Notching (4890)
Date of Event 08/26/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient presents with pain in her thigh post tha on the above listed date.There appears to be significant lysis around the femoral stem.This patient does not walk as she only transfers from a bed to a wheelchair.Stem is easily removed as the cement was only adhered to the stem proximally.Cement distal to the lytic activity was well fixed and left in place.Massive bone loss was noted both laterally and anteriorly.The liner was exchanged and a new stem and head were cemented with no delay nor patient harm.Cement pressurization was quite difficult with the missing bone but the new implants appeared to be well fixed and rom was deemed appropriate by the surgeon.Distal centralizer was unable to be removed.It was reported that patient was revised due to femoral stem loosening at unknown interface.Cement manufacturer was unknown.Doi: on (b)(6) 2021, dor: on (b)(6) 2026, affected side: left hip.
 
Additional Manufacturer Narrative
Product complaint#: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUMMIT
Common Device Name
Summit Hip Stem : Hip Femoral Stem
Manufacturer (Section D)
DePuy Orthopaedics Inc US
700 orthopaedic drive
warsaw IN 46581 -098
Manufacturer (Section G)
JTE Warsaw Mfg Site
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
+1(908)386-3534
MDR Report Key50790737
Report Number1818910-2026-15943
Device Sequence Number21475507
Product Code JDI
UDI-Device Identifier10603295060222
UDI-Public(01)10603295060222
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K170339
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number157013090
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 08/26/2026
Supplement Date Received by Manufacturer09/16/2026
Initial Report FDA Received Date09/16/2026
Supplement Report FDA Received Date09/17/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age82 YR
Patient SexFemale
-
-