Model Number M635TU90050
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Thrombosis/Thrombus (4440)
Date of Event 09/14/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported a thrombus occurred.A left atrial appendage (laa) closure procedure was being performed.A watchman trusteer access system (was) and a watchman laa closure device & delivery system (wds) were selected for use.The physician advanced the trusteer sheath and pigtail catheter into the appendage.An activated clotting time (act) measurement of 139 seconds was obtained.The physician paused to check act, this time measuring 143 seconds, and another 10,000 units of heparin was administered.The pigtail was removed from the trusteer sheath.As the pigtail catheter was removed from the patient's body, clots were noted on the pigtail catheter.The physician replaced the old trusteer for a new one.Act was measured and it was over 400 seconds.Physician proceeded with the implant procedure and a closure device was successfully implanted in the laa.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WATCHMAN TruSteer Access System
Common Device Name
Catheter, percutaneous
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
ballybrit business park
galway H91 Y 868
EI   H91 Y868
Manufacturer Contact
mary crawford
4100 hamline ave n
arden hills, MN 55112
+1(651)440-0979
MDR Report Key50791024
Report Number2124215-2026-49453
Device Sequence Number21186185
Product Code DQY
UDI-Device Identifier00191506022310
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K240018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberM635TU90050
Device Catalogue NumberM635TU90050
Device Lot Number0039757196
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/14/2026
Initial Report FDA Received Date09/16/2026
Date Device Manufactured07/08/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient Age79 YR
Patient SexFemale
Patient Weight138 LB
-
-