• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SynCardia Systems, LLC SynCardia Companion 2 Driver; External Pneumatic Driver

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SynCardia Systems, LLC SynCardia Companion 2 Driver; External Pneumatic Driver Back to Search Results
Catalog Number P/N 397002-001
Medical Device Problem Code Battery Problem (2885)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/13/2025
Type of Reportable Event Malfunction
Event or Problem Description
Per hospital staff, the companion 2 driver did not recognize external batteries at all.They tried cleaning the driver and tried multiple batteries that worked in other drivers without issue.Driver failed system check prior to being placed on a patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SynCardia Companion 2 Driver
Common Device Name
External Pneumatic Driver
Manufacturer (Section D)
SynCardia Systems, LLC
1992 e silverlake road
tucson AZ 85713
Manufacturer (Section G)
SynCardia Systems, LLC
1992 e silverlake road
tucson AZ 85713
Manufacturer Contact
michael larsen
1992 e silverlake road
tucson, AZ 85713
+1(520)545-1234
MDR Report Key50791253
Report Number3003761017-2025-00109
Device Sequence Number9443114
Product Code LOZ
UDI-Device Identifier00858000003107
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P030011
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 09/15/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberP/N 397002-001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 05/13/2025
Initial Report FDA Received Date09/16/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-