| Model Number |
LifeVest WCD 5100 System |
| Medical Device Problem Code |
Key or Button Unresponsive/not Working (4063)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/18/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A us distributor reported that a monitor's response buttons were not functioning.
|
| |
|
Additional Manufacturer Narrative
|
|
During investigation of the monitor on a reportable malfunction was found.The monitors response button was difficult to activate due to contaminated.The cause of the inability to activate the monitor is the contaminated response button.The root cause of the contaminated response button is liquid ingress of an unknown contaminant.Lifevest patient training materials have been updated as a reminder not to expose lifevest electronic components to liquid.No adverse event resulted from the damaged monitor.
|
| |
|