Mentor Texas MENTOR Smooth Round Moderate Plus Profile; Prosthesis, breast, inflatable, internal, saline Back to Search Results
Catalog Number 3502375
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Deformity/ Disfigurement (2360)
Date of Event 08/26/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient implanted with a 375cc mentor smooth round moderate plus profile saline breast prosthesis experienced a deflation on the left side post-operatively.As a result, the patient underwent explantation on (b)(6) 2026.
 
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: deflation.Section d6b, date of explant: (b)(6) 2026.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR Smooth Round Moderate Plus Profile
Common Device Name
Prosthesis, breast, inflatable, internal, saline
Manufacturer (Section D)
Mentor Texas
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
Mentor Texas
3041 skyway circle north
irving TX 75038
Manufacturer Contact
henry boden
3041 skyway circle north
irving, TX 75038
+1(949)788-2078
MDR Report Key50791428
Report Number1645337-2026-10329
Device Sequence Number21225105
Product Code FWM
UDI-Device Identifier00081317001423
UDI-Public(01)00081317001423(17)231107(10)9393617
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P990075
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/07/2023
Device Catalogue Number3502375
Device Lot Number9393617
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 08/27/2026
Initial Report FDA Received Date09/16/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/08/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age50 YR
Patient SexFemale
Patient RaceBlack Or African American
-
-