Mentor Texas MENTOR MemoryGel Breast Implant; Prosthesis, breast, noninflatable, internal, silicone gel-filled
|
Back to Search Results |
|
| Catalog Number |
3505501BC |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/07/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Device evaluation summary:\
\
the product was returned to mentor for evaluation, where a thorough investigation was conducted.\
\
during visual evaluation no apparent damage or visual anomalies were observed on the returned device.\
\
lymphadenopathy or palpable lymph nodes may be the result of bacterial or viral infection, neoplasm (primary or metastatic), or as the result of an idiopathic inflammatory process or foreign body reaction.There have been reports regarding movements of silicone gel material to lymph nodes even with or without evident of rupture leading to lymphadenopathy.A number of epidemiology studies have evaluated large populations of women with breast implants from a variety of manufacturers and implant models.The studies do not, taken together, support an association of breast implants and a diagnosed rheumatic disease.Other than one study, these studies do not distinguish whether the women had ruptured or intact implants.Because there were no supporting pathology reports or physician consultation summaries forwarded to product evaluation, we are unable to confirm the presence of or the possible etiology of the reported lymphadenopathy.Lymphadenopathy is a known complication associated with this surgery and is referenced in our current product insert data sheet.\
\
as part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria: since no malfunction was observed during the investigation.No corrective and preventive action (capa) is required now.\
\
additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.\
\
manufacturer¿s reference number:\
(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Mentor became aware the patient did not experience a rupture.Coding has been updated to indicate the patient's implant visibility/palpability through the skin.\
\
h6 health effect - clinical code e2402: breast implant palpability/visibility.\
\
manufacturer¿s reference number:\
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient implanted with a 550cc mentor memorygel breast implant experienced a rupture on the left side post-operatively.As a result, the patient underwent explantation on (b)(6) 2026.
|
| |
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.\
\
a manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.\
\
reason for device explant and/or reoperation: rupture.\
\
mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.\
\
manufacturer¿s reference number:\
(b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|