The customer reported the ophthalmic eye wash solution was leaking in the pre made kits, the papers were obviously stained, the eye wash bottles were stained and had a lower volume, and the cases have no shipping damage.Medline reviewed finished good production records, finished good pack build, trending, current stock, and a physical sample of four packs; the stock check found no defects and the production record had no line-down situations, runs without approvals, deviations, or non-conformance reports that might have had an impact on this issue.Sample evaluation found visible water damage on multiple components, bottles not full of the eye wash solution, liquid damage to the outside of the pack, and some eye washes had a loose heat seal on the top; this confirmed the issue.Medline determined this to be a vendor-product issue, noting the item was received pre-sealed from the component supplier and there is no additional manipulation during the manufacturing process, and provided the complaint to the vendor for further evaluation while supplier quality will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|