• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ICU Medical, Inc. CADD Solis HPCA Pump; Pump, infusion, pca

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ICU Medical, Inc. CADD Solis HPCA Pump; Pump, infusion, pca Back to Search Results
Model Number 2110
Medical Device Problem Code Device Alarm System (1012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/01/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during power up the pump experienced error code 41100 is associated with an adc safety signal issue.The pump was connected to a wireless communication module at the time of the fault.The errors only occurred after the communication modules were attached and power was then supplied via the ac adapter before being turned on.There was no patient involvement.
 
Additional Manufacturer Narrative
B3 ¿ date of event: the exact date of the event is unknown; therefore, the first day based on the icu medical awareness date was entered as the event date.The suspected device was not returned for evaluation.Therefore, visual inspection, functional testing, and event history log review could not be performed.A review of the available service history identified no previous related repairs within the last 12 months.The customer complaint pump displayed error code ¿41100¿ could not be confirmed.Based on the information provided and the absence of the device for evaluation, a probable cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD Solis HPCA Pump
Common Device Name
Pump, infusion, pca
Manufacturer (Section D)
ICU Medical, Inc.
6000 nathan lane north
minneapolis MN 55442
Manufacturer Contact
reed covert
600 n. field dr.
lake forest, IL 60045
+1(224)706-2300
MDR Report Key50792203
Report Number3012307300-2026-08736
Device Sequence Number9415577
Product Code MEA
UDI-Device Identifier15019517154290
UDI-Public(01)15019517154290(11)251006
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K170982
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number2110
Device Catalogue Number21-2111-0402-51
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/21/2026
Initial Report FDA Received Date09/16/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/06/2025
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-