| Catalog Number |
07028121190 |
| Medical Device Problem Codes |
High Test Results (2457); High Readings (2459)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/25/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
We received an allegation about discrepant results for 1 patient sample tested with elecsys vitamin b12 ii assay on a cobas 8000 e 801 module.Initial result: 756.No questionable result was reported outside the laboratory, and the sample was repeated on a different module.Repeat result: 364.The unit of measurement was not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The cobas 8000 e 801 module serial number was (b)(6).The alarm trace showed multiple sample clot alarms.The customer performed an air purge, liquid flow cleaning, and qc, and they were within specifications.The instrument was performing within specifications.No further issues were reported after the intervention.The instrument maintenance steps resolved the issue.The cause of the event could not be determined, but it was consistent with either an instrument issue requiring maintenance or sample interference (hyperproteinemia causing protein gel formation in assay cups).
|
| |
|