| Catalog Number |
05167043190 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/20/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
There was an allegation of questionable tina-quant albumin gen.2 results for 1 patient sample on a cobas 8000 cobas c 701 module.Initial result: 80.7 mg/l
the sample was repeated on another module, and the results were 27.8 mg/l and 27.8 mg/l.The sample was repeated on the same module as the initial result, and the result was 31.2 mg/l.
|
| |
|
Additional Manufacturer Narrative
|
|
The analyzer's serial number is (b)(6).The field service representative washed the probe.The investigation is ongoing.
|
| |
|