| Model Number |
CF-HQ190I |
| Medical Device Problem Codes |
Communication or Transmission Problem (2896); Connection Problem (2900); Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/20/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported the communication error e216, e237 and e226 during a inspection for use.During the initial evaluation of the device it was observed, e217 (scope error) occurred and micro connector unit in scope connector is dirty.No patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection and the reported failure was confirmed.A root cause could not be identified.Based on the results of the investigation.The most probable cause of the complaint is cause traced to component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|