| Model Number |
2110 |
| Medical Device Problem Code |
Failure to Infuse (2340)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the pump experienced error code 41100.Error code 41100 is associated with an adc safety signal issue.This error condition is associated with a high-priority alarm and could potentially interrupt therapy if it occurs during patient use.The pump was connected to a wireless communication module at the time of the fault.The event occurred during setup, and the pump was not in use when the error was generated.There was no patient involvement or harm.
|
| |
|
Additional Manufacturer Narrative
|
|
B3 ¿ date of event: the exact date of the event is unknown; therefore, the first day based on the icu medical awareness date was entered as the event date.The suspect pump was not returned for evaluation.A review of the 12-month service history was conducted, and no relevant findings were identified.The customer-reported issue could not be confirmed, and the root cause of the reported error condition could not be determined because the device was not returned for analysis and evaluation.
|
| |
|