| Model Number |
BF-UC180F |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The videoscope, an olympus asset with no reported complaint, was returned by the customer.During evaluation of the device, a small chip in the pink rubber was observed on the bronchofiberscope.No patient involvement was associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
B3: olympus detected the issue during the device servicing; therefore, there is no information regarding the customer reported event date.The device was returned to olympus for inspection.Based on the results of the investigation, the probable cause of this complaint is expected or random component failure resulting from degradation problem, with no design or manufacturing issue identified.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|