| Model Number |
3DM-SYS-STD |
| Medical Device Problem Code |
Thermal Decomposition of Device (1071)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/17/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The field engineer reported that while onsite for an unrelated repair, a burning odor was detected, and a dtc error code was present.The system was immediately powered down.Upon investigation, a burned-out chip was identified on the dtc board.The board was replaced, which resoled the error.The system was confirmed to be functioning properly following repair.No patient or user impact was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device history record (dhr) review was unable to be conducted for the device at the time of this report.A follow-up report will be submitted in the event additional information is received.
|
| |
|