Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H6 component codes: most relevant component code is g07002 (appropriate term/code not available) to capture no findings available due to no product returned.Investigation summary : according to the information provided, it was reported that during the surgical procedure, when 2 lcp 4.5/5.0 drill guides were inserted, they fractured, leaving fragments in the plate.One of the fragments is extracted, but the other remains inside the plate.There was no impact to patient.¿
a photo of the product was provided to depuy synthes for evaluation, and an investigation was conducted in accordance with depuy synthes procedures.¿
visual inspection of the of the photo found the image quality is poor and no observations pertaining to the nature of the reported event could be identified.¿
the reported complaint was not confirmed based on the evaluation results.Although no product defect was identified, other circumstances during device use may have contributed to the reported condition; however, those circumstances could not be identified from the evidence provided.¿
it has been determined that no corrective or preventive actions are proposed.There is no indication that a design or manufacturing issue caused the reported condition.As part of depuy synthes quality processes, all devices are manufactured, inspected, and released in accordance with approved specifications.Additional monitoring for potential safety signals is conducted through complaint trend analysis as part of post-market safety surveillance activities.Device history review : part: 323.042-15.Lot: 50024p4.Manufacturing site: hägendorf
release to warehouse: 04-apr-2025.A dhr evaluation was performed for the finished device article #323.042-15 lot #50024p4, and no non-conformances were identified.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|