It was reported that "the issue is lint residue transferring from the sterile towels onto the sterile field." "as you can see, lint and thread residue are on the wires, stent-triever, and sterile gloves.".
|
|
The customer reported that they are repeatedly seeing lint come off towels during cases, and lint and thread residue are on the wires, stent-triever, and sterile gloves.Medline received photos and no physical sample was available; upon examination of the photos, loose blue pieces of lint were observed on a guidewire and gloves, the finished good pack build was reviewed, and no issues with the build were identified.Based on the complaint details and photo evaluation, the reported loose lint was confirmed.Due to the nature of all woven textiles linting is an inherent possibility, certain uses of a woven product may increase the likelihood of occurrence, and the division will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|