A complaint was received regarding the item being "stuck" to a customer's wound, and no medical intervention, serious injury, or follow-up care has been reported.No physical or photo samples of the item or complaint issue were received from the customer; the vendor provided a coa, the product lot passed all testing and inspection criteria, and trending was conducted over the past two years by all versions of cur47147rbz and defect type (wound pad issues).Based on the results of the investigation, medline was unable to confirm the reported issue or that the product failed to meet specifications.No need for further action such as initiation of capas, scars and/or other risk management activities was indicated at this time, and complaints are monitored via the sop-00032 trend analysis procedure.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|