| Model Number |
IO-401 |
| Medical Device Problem Code |
Insufficient Flow or Under Infusion (2182)
|
| Health Effect - Clinical Codes |
Dyspnea (1816); Low Oxygen Saturation (2477)
|
| Date of Event |
08/19/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient's unit was displaying a replace columns message and it was not outputting enough oxygen.As a result, the patient had a hard time breathing.A replacement unit was sent to the patient.The inogen team attempted to contact the patient for further information three times with no response.
|
| |
|
Additional Manufacturer Narrative
|
|
B3 date of event is estimated.The unit was returned for evaluation on 31-aug-2026.The unit was received with a reported oxygen error, confirmed with the contaminated zeolite.Incoming internal 02 measured 72.5% and external 02 measured 72.8%, both below specification.The issue was identified as high psa (15) caused by zeolite contamination from moisture absorption.Also mount plate damaged due to compressor vibration.The uip & lower housing was replaced due to damaged.
|
| |
|