(b)(4).Date sent 10/7/2026
d4 batch # 827d89.Investigation summary
the product was returned for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that the cdh31p device was received with no apparent damage and the anvil returned inside a plastic bag.The washer was present, uncut and there were staples present in the device.The tissue compression scale did not backlight turn on when the test battery was inserted and the battery was difficult to insert and remove.The device was disassembled to verify the condition of the internal components and the bulkhead contact from the left side was noted to be damaged.In addition corrosion was noted on the motor.One possible cause for the condition of the motor may be the exposure of cleaning solution.No further functional testing could be performed due to the condition of the device.No conclusion could be reach on the cause of the bulkhead contact damage.It should be noted that when insert the battery pack, aligning the release tabs with the slots situated on the back of the device.Ensure that it is completely engaged; a clear click sound will confirm it¿s properly secured, and the backlight of the tissue compression scale will glow.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 827d89, and no non-conformances related to the reported complaint condition were identified.
|