| Model Number |
12773-03 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Malposition of Device (2616)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/25/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this was an arteriotomy closure of the calcified right common femoral artery using a prostyle device after an endovascular treatment procedure using an 8f sheath.Reportedly, after the plunger was depressed and withdrawn the suture was cut and the foot was retracted.The prostyle device was attempted to be removed; however, the device could not be removed.The prostyle device was able to be advanced slightly, the foot was repeatedly closed and opened, after which the device was able to be removed successfully.Under normal circumstances, the suture would engage the vessel wall and, once tied, achieve hemostasis; however, for an unknown reason, the suture did not capture the vessel wall and was instead pulled out without securing the tissue.Hemostasis was achieved with a non-abbott device.There were no reported adverse patient effects.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.Production record and corrective and preventative actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a query of the complaint handling database for the reported lot revealed there is no indication of a lot specific issue.Based on the reported information, the investigation determined that the reported issues could not be determined.Factors that may contribute to suture malposition include but not limited to under insertion that the vessel was not properly captured during suture deployment or interaction with the patient anatomy or friable femoral vessel contributed to the reported difficulty.Factors that contribute to difficult to remove vessel, include, but not limited to, device interaction with patient anatomy, tissue or suture caught in foot.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
| |
|
Search Alerts/Recalls
|
|
|