Maquet Cardiovascular LLC VasoView Hemopro 2; Electrosurgical, Cutting & Coagulation & Accessories Back to Search Results
Model Number VH-4000
Medical Device Problem Codes Peeled/Delaminated (1454); Material Twisted/Bent (2981)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 08/25/2026
Type of Reportable Event Serious Injury
Event or Problem Description
The hospital reported that vasoview hemopro 2 used in chh had split jaws.It happened on both side¿s, one side completely and the tip looked sharp.The surgeon did three length of vein and was about to finish last length of vein when it happened.Another device was used to complete the case and a redivac drain was inserted.Based on the provided photos the heating wire is twisted/bent and the insulation gap is visible.
 
Additional Manufacturer Narrative
(b)(4) the device has not yet been returned to maquet cardiac surgery for evaluation.We are following up with the customer for the return of the device.A supplemental report will be submitted if the device is received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VasoView Hemopro 2
Common Device Name
Electrosurgical, Cutting & Coagulation & Accessories
Manufacturer (Section D)
Maquet Cardiovascular LLC
45 barbour pond drive
wayne NJ 07470
Manufacturer (Section G)
Maquet Cardiovascular LLC
45 barbour pond drive
wayne NJ 07470
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 07470
MDR Report Key50794824
Report Number2242352-2026-0000808
Device Sequence Number13389439
Product Code GEI
UDI-Device Identifier00607567700406
UDI-Public(01)00607567700406
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K101274
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse Practitioner
Type of Report Initial
Report Date (Section B) 09/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Operator of Device Health Professional
Device Model NumberVH-4000
Device Catalogue NumberVH-4000
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/25/2026
Initial Report FDA Received Date09/16/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
Unknown
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-