| Model Number |
VH-4000 |
| Medical Device Problem Codes |
Peeled/Delaminated (1454); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
08/25/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The hospital reported that vasoview hemopro 2 used in chh had split jaws.It happened on both side¿s, one side completely and the tip looked sharp.The surgeon did three length of vein and was about to finish last length of vein when it happened.Another device was used to complete the case and a redivac drain was inserted.Based on the provided photos the heating wire is twisted/bent and the insulation gap is visible.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4)
the device has not yet been returned to maquet cardiac surgery for evaluation.We are following up with the customer for the return of the device.A supplemental report will be submitted if the device is received.
|
| |
|