• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; DBD-REMOVER,STAPLE,SKIN,STERILE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; DBD-REMOVER,STAPLE,SKIN,STERILE Back to Search Results
Catalog Number DYNJ04058
Medical Device Problem Code Loose or Intermittent Connection (1371)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/01/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the staple remover is not properly removing staples.The pin that connects the piece is loose.
 
Additional Manufacturer Narrative
The complaint reported a defective staple remover, stating that the pin that connects the piece is loose, and the issue did not occur during use and no injury occurred.A sample was received for investigation; medline received an unused stapler remover for sample review, reviewed finished good production records and finished good pack build, and functionally tested the staple remover by inserting staples into a skin pad and attempting to remove them.The staple remover handles were loose and did not hold shape, the pin appeared loose, and the staples did not remove in one piece, if at all as the remover did not grip the staples properly; the reported problem was confirmed.This component is not altered before being placed into the kit and the root cause is considered to be a component manufacturing product issue relating to the components manufacturing process; the division will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
DBD-REMOVER,STAPLE,SKIN,STERILE
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
venkata chittajallu
three lakes drive
northfield, IL 60093--275
MDR Report Key50797298
Report Number1417592-2026-43160
Device Sequence Number16079858
Product Code GDT
UDI-Device Identifier10080196846924
UDI-Public10080196846924
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/15/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberDYNJ04058
Device Lot Number23KBT405
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/01/2025
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/01/2025
Initial Report FDA Received Date09/17/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-