Model Number LSP112V
Medical Device Problem Codes Stretched (1601); Use of Device Problem (1670); Low impedance (2285)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the patient presented in clinic for initial implant procedure on (b)(6) 2026.Post fixation during procedure, the right ventricular (rv) leadless pacemaker (lp) exhibited no increase in pacing impedance and little to none current of injury.Repositioning was attempted, and the rvlp helix was noted to be sprung under fluoroscopy imaging.The rvlp was not implanted, and an alternate rvlp was implanted instead.There were no patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Aveir¿ Leadless Pacemaker, Right Ventricle
Common Device Name
Leadless Pacemaker
Manufacturer (Section D)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key50797300
Report Number2017865-2026-19724
Device Sequence Number18769484
Product Code PNJ
UDI-Device Identifier05415067034472
UDI-Public(01)05415067034472(10)S000109047(17)271027
Combination Product (Y/N)Y
Initial Reporter CountryAS
PMA/510(K) Number
P150035
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLSP112V
Device Lot NumberS000109047
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/01/2026
Initial Report FDA Received Date09/17/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/27/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-