• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Cordis US Corp. MYNXGRIP; Device, Hemostasis, Vascular

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Cordis US Corp. MYNXGRIP; Device, Hemostasis, Vascular Back to Search Results
Catalog Number MX5021
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/20/2026
Type of Reportable Event Malfunction
Event or Problem Description
As reported, the plug of a mynx grip vascular closure device (vcd) could not be released normally.There was no reported patient injury.The user shuttled down completely.There was no significant resistance when shuttling down, and no unusual force was applied when retracting the sheath.The advancer tube was engaged or visible.No damage was found on the device or device packaging prior to opening.The device was stored and prepared in accordance with the instructions for use (ifu).There was no reported patient injury.The mynx vascular closure device (vcd) was used in a diagnostic procedure with a retrograde approach.The deployer was mynx certified.The sheath french size was 6f.The reported vascular sheath introducer insertion angle was 30-45 degrees.The vessel was the femoral artery and was greater than 5 mm in diameter.There was no vessel tortuosity.The reported stick location was behind the femoral artery.There was no reported peripheral vascular disease (pvd) or calcium in the vicinity of the puncture site.Hemostasis was achieved by manual compression for 20 minutes.The device will be returned for evaluation.
 
Additional Manufacturer Narrative
Additional information is pending and will be submitted within 30 days upon receipt.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MYNXGRIP
Common Device Name
Device, Hemostasis, Vascular
Manufacturer (Section D)
Cordis US Corp.
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer (Section G)
Cordis US Corp.
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer Contact
karla castro
5452 betsy ross drive
santa clara, CA 95054
+1(786)313-8372
MDR Report Key50797364
Report Number3004939290-2026-01297
Device Sequence Number21213533
Product Code MGB
UDI-Device Identifier10862028000403
UDI-Public(01)10862028000403(17)280415(10)F2610506
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P040044
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberMX5021
Device Lot NumberF2610506
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/31/2026
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/20/2026
Initial Report FDA Received Date09/17/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/15/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-