| Model Number |
E142 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/26/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) and right ventricular (rv) lead recorded a shock lead impedance of 10 ohms during a hospital check.Technical services (ts) was consulted and noted this was likely due to electromagnetic interference (emi) given the hospital environment.It was noted that the local company representative repeated the test and a measurement of 90 ohms was obtained.No further out of range measurements were noted.This icd and rv lead remain in service.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|