| Model Number |
470230-15 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/18/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted surgical procedure, a medium large clip applier instrument had an exposed wire.The arm was broken and nonfunctional.The procedure was completed with no reported injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) received the da vinci product to perform failure analysis.The medium large clip applier instrument was analyzed and found the instrument was tested on an in-house system.The instrument passed the recognition and engagement tests.The instrument moved intuitively with a full range of motion in all directions.The grip tips opened and closed properly.The clip was unable to be loaded on the grips successfully.The instrument failed the clip test.The instrument was found to have bent grips.No other product issue was found.The instrument was found to have one (1) severely bent grip, causing misalignment of the grip-tips.There was a.73mm offset at the tips.A clip was able to be loaded into the clip groove of the grip-tips.The clip applier failed the clip tests.Instrument was installed on an in-house system and driven.The clip applier instrument failed the clip test due to the grip tips being severe.The clip was unable to be installed on the grips properly, causing clip test failure.The complaint regarding the instrument had an exposed wire and a broken arm was not confirmed by failure analysis.The probable root cause can be attributed to damage during either use or reprocessing causing improper grip alignment, preventing proper clip formation.
|
| |
|
Search Alerts/Recalls
|