| Model Number |
0998-00-0800-45 |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by the customer that during use, the cardiosave intra-aortic balloon pump (iabp) blood pressure was too low.No patient involvement.No harm was reported.A getinge field service engineer (fse) was dispatched to evaluate the unit.The fse reported that all iabp tests were normal, and balloon inflation and deflation were normal.Product passed all functional and safety tests per manufacturer specifications.The device is ready for clinical use.
|
| |
|
Additional Manufacturer Narrative
|
|
Event site postal code: (b)(6).Upon completion of the investigation into this event a follow up report will be submitted.
|
| |
|