• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medtronic Xomed Inc. NIM® 3.0 INTERFACE; STIMULATOR, NERVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medtronic Xomed Inc. NIM® 3.0 INTERFACE; STIMULATOR, NERVE Back to Search Results
Model Number 8253200
Medical Device Problem Code Inappropriate Audible Prompt/Feedback (2280)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/24/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that pre op there was a great interference occurred from the patient interface.There was no patient involved.
 
Additional Manufacturer Narrative
H3: analysis found that when the pi is connected to the standard console and the system enters the monitoring screen, significant interference is observed, resulting in an abnormal waveform.There was cable failure.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIM® 3.0 INTERFACE
Common Device Name
STIMULATOR, NERVE
Manufacturer (Section D)
Medtronic Xomed Inc.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer (Section G)
Medtronic Xomed Inc.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer Contact
justin ellis
6743 southpoint drive north
jacksonville, FL 32216
+1(763)526-5677
MDR Report Key50798285
Report Number1045254-2026-01282
Device Sequence Number10610872
Product Code ETN
UDI-Device Identifier00643169283275
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K083124
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8253200
Device Catalogue Number8253200
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/26/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/25/2026
Initial Report FDA Received Date09/17/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/25/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-