The physical device was not returned for investigation.Cloud data for the period of (b)(6)2026 was downloaded and reviewed.Patient history buffer data indicated that the system was operating in both automated and manual modes.While in automated mode, the algorithm functioned as intended to adapt insulin delivery in response to fluctuations in estimated glucose values (egvs).As egvs increased, the algorithm increased microbolus delivery to correct egvs towards the user's setpoint.As egvs began to plateau and decrease, the algorithm appropriately suspended microbolus delivery until egvs began to increase once again.While operating in manual mode, the system correctly delivered insulin at the user's set basal segments of 3 and 18 pulses/hour respectively.Cloud records indicate that automated delivery restriction alerts were generated at 10:20 on(b)(6)2026, 14:00 on (b)(6)2026, 10:36 on(b)(6)2026, and 17:31 on (b)(6)2026.In each instance of an automated delivery restriction alert, the system was observed to have been delivering at the maximum adaptive rate for an extended period of time.As the alerts generated, the system correctly transitioned into automated mode: limited, delivered insulin at the user's safe basal rate.As the alerts were manually acknowledged, the system then transitioned into manual mode as expected.Missing egv alerts were prompted at 06:55 and repeated at 07:56 on (b)(6)2026 after a string of 1 and 5 egvs, indicating that the sensor was not active and the sensor was out of calibration, respectively.After a pod change at approximately 08:57, the pod and sensor resumed normal connection, with the pod receiving expected egvs.The system was correctly delivering insulin at the user's safe basal rate while the missing egv alert was active.After the automated delivery restriction alert from 17:31 on (b)(6)2026 was acknowledged, the system remained in manual mode for the duration of system use.The final active pod observed in the cloud data (sequence number 200803) was deactivated at 23:07 on (b)(6)2026.The last available egv reading was 103 mg/dl at approximately 23:07, concurrent with the last observed pod deactivation.Review of all bolus records for the period of cloud data review confirm that all programmed boluses were delivered to completion.The pulses delivered count was observed to have incremented as intended in response to system inputs, confirming that the user's pods were delivered the correct amount of insulin as programmed.No problems were found that would indicate that the system was failing to administer insulin as intended.Lot release records were reviewed and the product lot met all acceptance criteria.Locked down smartphone: data not provided,
omnipod software app version: 3.1.6,
operating system: n5004l-am-q-mv01602-06-01.06,
hardware: n5004l,
cgm sensor type: data not provided.Please note, the device identifiers are captured as reported by the complainant and may not align with the device configuration reported in this section as this data is pulled from our cloud based on the reported date of event.
|