| Model Number |
0998-00-0800-55 |
| Medical Device Problem Codes |
Inflation Problem (1310); Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/25/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by the user that prior to use, the cardiosave intra-aortic balloon pump (iabp) did not inflate balloon.There was no patient involvement.No harm was reported.A getinge field service engineer (fse) was dispatched to evaluate the unit.The fse reported that unit had power up test fail# 15.Pim was disassembled and checked.Unit was calibrated.Product passed all functional and safety tests per manufacturer specifications.
|
| |
|
Additional Manufacturer Narrative
|
|
Event site name: (b)(6).Upon completion of the investigation into this event a follow up report will be submitted.
|
| |
|