| Model Number |
0998-00-3023-55 |
| Medical Device Problem Codes |
Electrical /Electronic Property Problem (1198); Failure to Read Input Signal (1581)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/28/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a routine check the cs300 intra-aortic balloon pump (iabp) electrical test fails 39 and electrical test fails 65 occurred.No patient involved.No harm was reported.The issue discovered by the customer.A getinge field service engineer (fse) was dispatched to evaluate the unit.The fse confirm that the eletrical faile codes occurred.Board is suspected to be defective.After the replacement of main board, errors 39 and 65 no longer appeared.After further testing a machine exhibited a.P.Optical sensing module failure.Fiber optic sensor module (fos) is suspected to be defective and needs to be replaced.After replacement of fos module machine is working properly and was handed to user.
|
| |
|
Additional Manufacturer Narrative
|
|
Event site postal code: (b)(6).Upon completion of the investigation into this event a follow up report will be submitted.
|
| |
|