MEDTRONIC PUERTO RICO OPERATIONS CO. Percept; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
|
Back to Search Results |
|
| Model Number |
B35300 |
| Medical Device Problem Codes |
Improper or Incorrect Procedure or Method (2017); Battery Problem (2885); Charging Problem (2892)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/19/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the patient normally only needs to recharge every 3-4 days, but they let the battery hit 0% on one point on accident and since then, it has been losing 40% to 50% in a single day and they are having to recharge more.Issue started about 2.5 wks ago.The caller asked if the patient had changed any programs or groups, but the caller said probably not.The patient is in mexico city and does not have a treating neurologist there, rep stated they are unlikely to change settings.Reviewed with the caller to have the patient keep a diary of their charging interval, charging duration and start/end battery level.If the issue continues, redirect to a physician that can check the system and pull logs to check recharge history.
|
| |
|
Additional Manufacturer Narrative
|
|
B3: date is approximate.Month and year are confirmed valid.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|