• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; BED, LOW FULL ELECTRIC, BASIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; BED, LOW FULL ELECTRIC, BASIC Back to Search Results
Catalog Number MDR107003ELO
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/01/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "broken pully system.".
 
Additional Manufacturer Narrative
The customer reported a "broken pully system" for the bed.Medline reviewed photo evidence provided by the customer; a physical sample was not provided, so only a visual review of the defect could be conducted and a functional or dimensional inspection could not be executed via a photo.The investigation results were confirmed: the product appeared to be in used condition, and the right side leg of the footboard appeared to be extended while the left had contracted, showing that the right side pulley system failed to pull the leg up.The documented cause was "dc: unknown responsible party - cannot determine cause w/received sample"; no trends were identified, no product changes were identified that could have contributed to the alleged product failure, and the division will monitor complaints through quarterly trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
BED, LOW FULL ELECTRIC, BASIC
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
kumar vigrahala
three lakes drive
northfield, IL 60093--275
MDR Report Key50801090
Report Number1417592-2026-43190
Device Sequence Number8653447
Product Code LLI
UDI-Device Identifier10080196546992
UDI-Public10080196546992
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMDR107003ELO
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/01/2025
Initial Report FDA Received Date09/17/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-