| Catalog Number |
010002622 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/17/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that while using the g7 liner provisional, size c (32 mm, neutral), the plastic portion of the trial liner broke during removal after completion of the trial reduction.No patient harm or impact was reported.It was confirmed that no further information could be provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).G2: foreign - japan.Visual examination of the returned product identified signs of use as well as wear and tear.There are gouges and knicks on the outside diameter and rim of the liner.Both tabs near the through hole have fractured as shown in the photos.The provisional screw was also returned with the device and had some light scratches and knicks on the on the top.Verified outside diameter and that the liner is visually conforming to print.The device has an approximate field age of 7 years.Review of the device history records identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor trends.
|
| |
|