| Model Number |
9-ASD-014 |
| Medical Device Problem Code |
Difficult to Fold, Unfold or Collapse (1254)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/20/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026, 14mm amplatzer septal occluder was chosen for implant using an unknown delivery system.During the procedure, it was noted that the device conformed to cobra shape once deployed in the left atrium.The deformed device was never released from the delivery cable while inside the patient.There was no angulation or kink noted in the delivery system.There was no interaction with cardiac structures during deployment.The physician doctor recaptured the device to check in water again; it took the same aspect.A new 14mm amplatzer septal occluder was used as a replacement device to complete the procedure.The patient remained hemodynamically stable throughout the procedure.There was no clinically significant delay in procedure and no adverse patient effects reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of device deformity was reported.A returned device assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all defined manufacturing specifications.Additionally, the lot was reviewed for similar complaints, and there is no indication of a lot-specific product issue.Based on the information received, the cause of the reported incident could not be conclusively determined.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
|
| |
|
Search Alerts/Recalls
|