• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Maquet Cardiovascular LLC HST III System (3.8MM); Clamp, Vascular

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Maquet Cardiovascular LLC HST III System (3.8MM); Clamp, Vascular Back to Search Results
Model Number HSK-3038
Medical Device Problem Codes Fitting Problem (2183); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/21/2026
Type of Reportable Event Malfunction
Event or Problem Description
The hospital reported that there was an attempt to load a hst iii system (3.8mm) into a delivery device using the ifu method.However, the seal became stuck on the loader, preventing normal loading, and the delivery device handle detached.Determined to be a product defect, a new product was used, and the surgery was successfully completed.The patient's condition is stable.There was a procedural delay of approximately 7¿8 minutes.
 
Additional Manufacturer Narrative
(b)(4).Event site name exceeded character limitations: (b)(6) hospital.The device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HST III System (3.8MM)
Common Device Name
Clamp, Vascular
Manufacturer (Section D)
Maquet Cardiovascular LLC
45 barbour pond drive
wayne NJ 07470
Manufacturer (Section G)
Maquet Cardiovascular LLC
45 barbour pond drive
wayne NJ 07470
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 07470
MDR Report Key50801463
Report Number2242352-2026-0000811
Device Sequence Number10590197
Product Code DXC
UDI-Device Identifier00607567700314
UDI-Public(01)00607567700314(10)3000500748(17)260902
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K130382
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 09/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/02/2026
Device Model NumberHSK-3038
Device Catalogue NumberHSK-3038
Device Lot Number3000500748
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/28/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/24/2026
Initial Report FDA Received Date09/17/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/02/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
Unknown
Patient SexUnknown
-
-