• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Abbott Medical CARDIOMEMS PATIENT ELECTRONIC SYSTEM; System, Hemodynamic, Implantable

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Abbott Medical CARDIOMEMS PATIENT ELECTRONIC SYSTEM; System, Hemodynamic, Implantable Back to Search Results
Model Number CM1100
Medical Device Problem Code Intermittent Loss of Power (4016)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/27/2026
Type of Reportable Event Malfunction
Event or Problem Description
The investigation of the device revealed a finding of frayed and exposed wires on the power extension cable of the device.The patient electronic unit was replaced and there were no adverse consequences reported by the patient.
 
Additional Manufacturer Narrative
Fa-q323-hf-4 cardiomems pes right angle power extension cable failure notice issued by abbott on 04 oct 2023.One cardiomems patient electronic system was received for evaluation.External visual inspection of returned material revealed exposed and frayed wires on the right ang ext cable.No additional physical damage was observed.The backplate of the device was removed and a standard power supply was connected.The unit was powered on with no issues observed.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The review determined that no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOMEMS PATIENT ELECTRONIC SYSTEM
Common Device Name
System, Hemodynamic, Implantable
Manufacturer (Section D)
Abbott Medical
387 nerem st nw suite 500
atlanta GA 30313
Manufacturer (Section G)
Abbott Medical
387 nerem st nw suite 500
atlanta GA 30313
Manufacturer Contact
fredric gold
168 middlesex turnpike
burlington, MA 01803
+1(651)756-5400
MDR Report Key50801505
Report Number3004936110-2026-01621
Device Sequence Number10610444
Product Code MOM
UDI-Device Identifier05414734509800
UDI-Public(01)05414734509800(10)10391418
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
P100045
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Remedial Action Notification
Type of Report Initial
Report Date (Section B) 09/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberCM1100
Device Catalogue NumberCM1100
Device Lot Number10391418
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/02/2026
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/10/2026
Initial Report FDA Received Date09/17/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/18/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
FA-Q323-HF-4
Patient Sequence Number1
Patient Age72 YR
Patient SexMale
-
-