Datascope Corp. CS300 Intra-Aortic Balloon Pump, English, 110V; System, Balloon, Intra-Aortic And Control
|
Back to Search Results |
|
| Model Number |
0998-00-3023-53 |
| Medical Device Problem Code |
Signal Artifact/Noise (1036)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/28/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by customer that during use, the cs300 intra-aortic balloon pump (iabp) showing a poor ecg waveform.Patient involved.No harm was reported.Users swapped out console to continue treatment without issue.A getinge field service engineer (fse) was dispatched to evaluate the unit.According to the fse users surfaced subject unit showing a ¿poor waveform¿ while on patient.The ¿poor waveform¿ appears to relate to the ecg and/or blood pressure waverforms, however unclear.Other than that, device was fully functional.Users swapped out console to continue treatment without issue.Fse was unable to replicate user complaint but found the ecg trunk cable and leads with significant ruptures on insulation jackets, hence likely related to the ¿poor waveform¿ complaint.Unit calibrated and passed all functional and safety tests per factory specifications.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon completion of the investigation into this event a follow up report will be submitted.
|
| |
|