Datascope Corp. CS300 Intra-Aortic Balloon Pump, English, 110V; System, Balloon, Intra-Aortic And Control
|
Back to Search Results |
|
| Model Number |
0998-00-3023-53 |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/28/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, the cs300 intra-aortic balloon pump (iabp) screen is not working.It is unknown when malfunction occurred.No patient involved.No harm was reported.Failure observed by end user screen.Tested cs300 and observed on power up distorted screen and vertically not stable.Isolated failure to video cable, returned part.Ordered display conversion kit and installed on pump.Unit passed all functional and safety tests per factory specifications, returned to customer.
|
| |
|
Additional Manufacturer Narrative
|
|
Event site name: (b)(6).\
\
upon completion of the investigation into this event a follow up report will be submitted.
|
| |
|