• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Datascope Corp. CS100 Intra-Aortic Balloon Pump, English, 220V; System, Balloon, Intra-Aortic And Control

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Datascope Corp. CS100 Intra-Aortic Balloon Pump, English, 220V; System, Balloon, Intra-Aortic And Control Back to Search Results
Model Number 0998-00-3013-55
Medical Device Problem Codes Battery Problem (2885); Power Problem (3010)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/28/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported by customer that during a routine check up, the cs100 intra-aortic balloon pump (iabp) did not turn on when plugged into an ac power outlet.The was no patient involved.No harm was reported.A getinge field service engineer (fse) was dispatched to evaluate the unit.The fse found that the ac power cord was damaged, the repair was conducted.The device was tested and found to be functional but the safety disc battery ome parts were worn out and past their lifespan.Safety disk is worn out and has reached the end of its lifespan.The battery for the iabp is worn out (expired) and the device cannot be used with the battery.
 
Additional Manufacturer Narrative
Event site telephone: (b)(6) upon completion of the investigation into this event a follow up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS100 Intra-Aortic Balloon Pump, English, 220V
Common Device Name
System, Balloon, Intra-Aortic And Control
Manufacturer (Section D)
Datascope Corp.
1300 macarthur blvd
mahwah NJ 07430
Manufacturer (Section G)
Datascope Corp.
1300 macarthur blvd
mahwah NJ 07430
Manufacturer Contact
ewelina chorlowska
1300 macarthur blvd
mahwah, NJ 07430
MDR Report Key50801566
Report Number2249723-2026-0004329
Device Sequence Number8596958
Product Code DSP
UDI-Device Identifier10607567107295
Combination Product (Y/N)N
Initial Reporter CountryTH
PMA/510(K) Number
K031636
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 09/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number0998-00-3013-55
Device Catalogue Number0998-00-3013-55
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/28/2026
Initial Report FDA Received Date09/17/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/17/2019
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-