| Model Number |
0998-00-3013-55 |
| Medical Device Problem Codes |
Battery Problem (2885); Power Problem (3010)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/28/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by customer that during a routine check up, the cs100 intra-aortic balloon pump (iabp) did not turn on when plugged into an ac power outlet.The was no patient involved.No harm was reported.A getinge field service engineer (fse) was dispatched to evaluate the unit.The fse found that the ac power cord was damaged, the repair was conducted.The device was tested and found to be functional but the safety disc battery ome parts were worn out and past their lifespan.Safety disk is worn out and has reached the end of its lifespan.The battery for the iabp is worn out (expired) and the device cannot be used with the battery.
|
| |
|
Additional Manufacturer Narrative
|
|
Event site telephone: (b)(6)
upon completion of the investigation into this event a follow up report will be submitted.
|
| |
|