(b)(4).Date sent 9/17/2026.D4 batch # unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this lot.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information was requested, and the following was obtained:
were there any adverse event/patient consequences in this case?
ans: nope.Did the same issue happened for both devices?
ans: yes.2.If not, did the device lock-out (no staples deployed and no cut line started)?
ans: yes.3.Or, did the device start to deploy staples and cut but could not be completed?
ans: never deployed stapler at all.4.Were any staples delivered into the tissue at all? ans: no.5.What was the appearance of the deployed staples (b-formed, malformed, goal post/legs straight)?
ans: no staplers deployed.6.Were there staples missing from the staple line (staples not present/visible on any portion of the staple line)?
ans: no.7.Did the device cut the tissue?
ans: no.8.If the device cut, was the cut line complete?
ans: no cutting.9.How was the issue addressed?
ans: not using staplers, surgeon handsaw.10.Please provide the source or title of external person providing answers to follow-up, e.G.Nurse, surgeon, etc.Ans: - surgeon.11.Device follow-up: is the product still available for return? if yes, please drop at the office or courier back the sample using gdex to j&j pinnacle office.Ans: - yes.A video was received.Any additional information obtained from the video evaluation will be included as part of the product investigation.
|