It was reported that when using the bd pyxis¿ medstation¿ es tower, the system became frozen on the blue screen, and the remote support service (rss) went offline.The customer rebooted the station, but issue persisted.The customer stated that this malfunction caused a delay while dispensing medications to the patient and the user managed to get medications from pharmacy.However, there were no adverse events or injuries reported based on this incident.
|
|
A review of the complaint history for sn: (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn: (b)(6) was performed from the date of manufacture, 29-apr-2026 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the station became frozen on the blue startup screen.A technical support specialist (tss) deployed a package to restart the remote support service (rss) service and rebooted the station, restoring remote connectivity.The station was reconfigured to automatically launch the carefusion (cfn) application, but an initialization device manager error was observed, and an onsite field service technician (fst) visit was scheduled.A field service engineer (fse) found the station powered off.The station was powered on successfully, booted normally, and allowed user login.Pharmacy technician performed inventory activities and confirmed operation, although intermittent slowness was noted.The tss done multiple reboots and application verification with no errors or issues were identified and the station returned to normal operation.The customer confirmed functionality, and the issue was resolved.The system functioned as intended after the technical support specialist and field service engineer investigated the device.
|