The customer reported that the needles have been coring the tops of medication vials, resulting in tiny plastic particles being released into the vial or the drawing syringe, and medication needed to be wasted and a new replacement pulled from the pyxis.Medline received 7bx standard hypodermic needles in new condition; per account details, the samples are used to draw medication, and the samples worked as intended.The complaint was not confirmed.According to the customer, "it seems we used the wrong needle and appears to be an internal issue"; hypodermic needles are not designed to penetrate vials, and the cause code was updated to reflect dc: traced to user - misuse/mishandling; the division has been informed and will continue to monitor trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|