Ethicon Endo-Surgery, LLC. ECHELON FLEX; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number PVE35A
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/10/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a bedside robotic surgery firing on an appendix, it was observed that the device did not fire nor cut.Procedure was delayed due to the reported event.There is no harm to the patient and procedure was completed with the robotic stapler.No additional information provided.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 9/17/2026.D4: batch #760d12.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one pve35a device was returned with no apparent damage and with a reload present.The reload was received partially fired 1/3.The manual override door out of position; the override lever was up which denotes that the knife was manually returned to home position.It should be noted after the manual override system is used the instrument is disabled and cannot be used for any subsequent firings.The bailout system was reset and then, the returned device and reload were tested for functionality in the straight position by resetting and reloading it into the device.The device achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the remaining staples met the staple form release criteria.The partially fired is consistent with an incomplete or interrupted cycle.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 760d12, and no non-conformances related to the reported complaint condition were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHELON FLEX
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
Ethicon Endo-Surgery, LLC.
475 calle c
guaynabo 00969
PR  00969
Manufacturer (Section G)
Ethicon Endo-Surgery, LLC.
475 calle c
guaynabo 00969
PR   00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 00969
PR   00969
+1(303)552-6892
MDR Report Key50801974
Report Number3005075853-2026-07631
Device Sequence Number21480108
Product Code GDW
UDI-Device Identifier10705036014584
UDI-Public(01)10705036014584(17)280830(10)760D12
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K141952
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberPVE35A
Device Lot Number760D12
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/28/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/31/2026
Initial Report FDA Received Date09/17/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/23/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
VASECR35 x2.
Patient SexUnknown
-
-