A.1.- a.5.Additional patient information was not provided.H.9.Livanova deutschland implemented a field safety notice for disinfection and cleaning of heater-cooler devices.The z number is z-2076/2081-2015.H11:
a legal lawsuit has been initiated, and no additional information is currently available.No clinical or device-related information was provided.Furthermore, the specific surgical procedure remains unknown, and no medical records or laboratory results have been made available to confirm the reported event.Review of complaints database identified other similar contamination events from the same hospital, both on the same and other devices.Device history record review of the involved device serial number manufactured in april 2017 was conducted and did not identify any deviations or non-conformities relevant to the reported issue.The service history review revealed that vacuum and sealing kit upgrade was performed on 26 february 2020.Based on the available information, source of patient contamination and involvement of the device remains unknown.Livanova has implemented a strategy to progressively decrease the probability of bacteria grow in the hc device by applying multiple measures implemented over the past few years through dedicated capa.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
|