The reported failure was that the pressure cuffs have air leaking and were not inflating/leaking during use.Medline received two handheld aneroid blood pressure monitor with d-ring cuffs in like-new condition, performed a visual inspection, confirmed that all parts were attached and no defects were noted, and inflated each cuff up to 300 mmhg five times according to the instructions for use with no issues experienced.The complaint was not confirmed, and no root cause was identified.The cuffs inflated/deflated as normal, the air release valve worked releasing the cuff¿s pressure as intended, and the samples worked as intended.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|