Micro Therapeutics, Inc. dba ev3 CONCERTO VERSA; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
|
Back to Search Results |
|
| Catalog Number |
CV-5-30 |
| Medical Device Problem Codes |
Entrapment of Device (1212); Premature Separation (4045)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/20/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during coil packing at the intended site for treatment of a right external femoral and profunda arteriovenous m alformation, a cv-5-30 versa coil prematurely detached from the pusher wire. two cv-6-20 coils were delivered without issue before the reported event. the artery measured 4mm, had minimal tortuosity, and abnormal vascular anatomy was reported.A 65cm 4fr non-medtronic (terumo glidecath) diagnostic catheter, a medtronic instant detacher, and a 45cm 6fr non-medtronic (terumo) sheath were used.The device was prepared according to the instructions for use, and a continuous heparinized saline flush was administered.The pusher delivery system was advanced more than 2 cm distally beyond the diagnostic catheter. the physician repositioned the coil and rotated the delivery pusher. the physician attempted to use the pusher wire to deliver the prematurely detached coil. the pusher wire got on other side of cv-5-30 coil and got stuck in the coil pack of the previously deployed cv-6-20 coils.After multiple different attempts to free the pusher wire were unsuccessful, the whole coil pack was removed from the patient and the case was started over. the issue was resolved on 20-aug-2026.Patient was treated with an abre stent, device was removed successfully with no injury to the patient.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Imageanalysis the customer returned a video clip.The video clip shows a concerto versa coil being advanced towards a coil that has already been deployed.The reported event cannot be confirmed.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|