| Catalog Number |
GLC80 |
| Medical Device Problem Code |
Unintended Compatibility (4047)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/11/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an unknown procedure the glc linear cutter broke from the anvil part.Unable to fire the cartridge.Was there any reported patient or user harm? no.Was there a significant delay? no.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 9/17/2026.D4 batch # 887d00.Investigation summary.The product was returned for evaluation.Visual inspection was conducted on the returned device.Visual examination of the returned glc80 device identified the knob in the forward position.The anvil was found broken and remained attached only by the anvil insert.Damage was also noted on the anvil shroud tab.No reload was available for evaluation.Due to the condition of the device no functional test was performed.No conclusion could be reached as to how the metal anvil became broken.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 887d00, and no non-conformances were identified.
|
| |
|