This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative:
investigation summary:
the product has not returned to depuy synthes, however a photo was provided for evaluation.¿
visual inspection of the returned photo found shaving particles into the arthroscopic space.No other anomalies could be observed.The overall complaint was confirmed as the observed condition of the aggressive 4.0mm 5pk would have contributed to the complained device issue.¿
based on the investigation findings, the potential cause can be traced to blade wear and degradation.As per instructions for use (ifu) excessive side loading may result in blade wear and degradation.Adequate suction is recommended to reduce wear and degradation of the device, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history:
a manufacturing record evaluation was performed for the finished device lot number, and no non-conformance was identified.
|