| Model Number |
5095 Surigical Table |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/18/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The user facility reported that during a patient procedure their 5095 surgical table ¿rolled away¿.No report of injury or procedure delay.
|
| |
|
Additional Manufacturer Narrative
|
|
A steris service technician arrived onsite following the reported event to inspect the unit and could not duplicate the reported event.The table was found to be operating properly and returned to service.No repairs were made.Based on the technician¿s discussion with the user facility biomed department, the cause of the reported event was likely attributed to user facility personnel inadvertently unlocking the table.The 5095 surgical table operator manual states, ¿warning personal injury hazard failure to engage the floor locks before placing a patient on or removing from the table or disengaging the floor locks while a patient is on the table during a surgical procedure could result in the table rolling unexpectedly during the procedure.¿
the technician counseled user facility personnel on the proper use and operation of the table, specifically keeping the floor locks engaged.No additional issues have been reported.
|
| |
|